StacksVerified U.S. regulatory reference

40 CFR Part 53, Table F-1

Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov
Performance testSpecificationsAcceptance criteria
§ 53.62 Full Wind Tunnel EvaluationSolid VOAG produced aerosol at 2 km/hr and 24 km/hrDp50 = 2.5 µm ±0.2 µm Numerical Analysis Results: 95% ≤Rc ≤105%.
§ 53.63 Wind Tunnel Inlet Aspiration TestLiquid VOAG produced aerosol at 2 km/hr and 24 km/hrRelative Aspiration: 95% ≤A ≤105%.
§ 53.64 Static Fractionator TestEvaluation of the fractionator under static conditionsDp50 = 2.5 µm ±0.2 µm Numerical Analysis Results: 95% ≤Rc ≤105%.
§ 53.65 Loading TestLoading of the clean candidate under laboratory conditionsAcceptance criteria as specified in the post-loading evaluation test (§ 53.62, § 53.63, or § 53.64).
§ 53.66 Volatility TestPolydisperse liquid aerosol produced by air nebulization of A.C.S. reagent grade glycerol, 99.5% minimum purityRegression Parameters Slope = 1 ±0.1, Intercept = 0 ±0.15 mg, r ≥0.97.

[72 FR 32209, June 12, 2007]